SHEN Xiao-pan, LI Zhi-ming, ZHAI Li-hua, DENG Hu, XU Jiang, WEI Guan-yi. Selective Ionization of Trace Plutonium from Uranium-Plutonium Mixture by Laser Resonance Ionization Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2013, 34(5): 269-272. DOI: 10.7538/zpxb.2013.34.05.0269
Citation: SHEN Xiao-pan, LI Zhi-ming, ZHAI Li-hua, DENG Hu, XU Jiang, WEI Guan-yi. Selective Ionization of Trace Plutonium from Uranium-Plutonium Mixture by Laser Resonance Ionization Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2013, 34(5): 269-272. DOI: 10.7538/zpxb.2013.34.05.0269

Selective Ionization of Trace Plutonium from Uranium-Plutonium Mixture by Laser Resonance Ionization Mass Spectrometry

More Information
  • Based on the laser resonance ionization mass spectrometer (LRIMS) developed in the laboratory, a preliminary method was established to determine plutonium isotope ratios. Through scanning the wavelength of ionizing laser, a three-color-three-photon resonance ionization scheme for plutonium with greater ionization section was obtained. A uranium plutonium mixture was prepared to be used as a simulant sample, and the plutonium isotope ratio in the sample was determined. Results show that the relative standard deviation of 240Pu/239Pu isotope ratio is 1.1%, and the element selectivity of plutonium to uranium is better than 5×106. This work indicates that LRIMS can effectively avoid the isobaric interference and strong-peak-tail interference from other elements.
  • [1]
    COOLEY J N, KUHN E, DONOHUE D L. Proceedings of IAEA symposium on international safeguards, IAEA-SM-351/182[R]. Vienna, Austria: IAEA, 13-17, Oct, 1997.
    [2]
    NISHIMURA H, MAGARA M, ODA T. Proceedings of IAEA symposium on international safeguards, IAEA-SM-351/98[R]. Vienna, Austria: IAEA, 18-22, Oct, 1997.
    [3]
    INKRET W C, EDURD D W, MILLER G, et al. Applications of thermal ionization mass spectrometry of the detection of 239Pu/240Pu intakes[J]. International Journal of Mass Spectrometry, 1998,178: 113-120.
    [4]
    MAGARA M, HANZAWA Y, ESAKA F, et al. Development of analytical techniques for ulta trace amounts of nuclear materials in environmental samples using ICP-MS for safeguards[J]. Applied Radiation and Isotopes, 2000,53: 87-90.
    [5]
    OUGHTON D H, SKIPPERUD L, FIFIELD L K, et al. Accelerator mass spectrometry measurement of 240Pu, 239Pu isotope ratio in Novaya Zemlya and Kara Sea sediments[J]. Applied Radiation and Isotopes, 2004,61: 249-253.
    [6]
    李力力, 李金英, 赵永刚, 等. 多接受电感耦合等离子体质谱法精确测量铀中痕量钚的同位素方法研究[J]. 质谱学报, 2009, 26(1): 6-10.
    LI Lili, LI Jinying, ZHAO Yonggang, et al. Studies on the precise measurement of isotope of trace plutonium in uranium by MC-ICP-MS[J]. Journal of Chinese Mass Spectrometry Society, 2009, 26(1): 6-10 (in Chinese).
    [7]
    周国庆, 朱凤蓉, 李梅, 等. 石墨探针进样电感耦合等离子质谱法测定强铀干扰下超微量240Pu/239Pu同位素比[J]. 分析化学,2011, 39(11): 1682-1688.ZHOU Guoqing, ZHU Fengrong, LI Mei, et al. Isotopic ratio of 240Pu/239Pu measurements under interference of uranium by inductively coupled plasma-mass spectrometry with direct sampling insertion[J]. Chinese Journal of Analytical Chemistry, 2011, 39(11): 1682-1688 (in Chinese).
    [8]
    PAYNE M G, DENG L. Application of resonance ionization mass spectrometry[J]. Rev Sci Instrument, 1994, 65(8): 2433-2459.
    [9]
    LETOKHOV V S. Laser photoionization spectro-scopy [M]. Lodon: Academic Press INC, 1987.
    [10]
    CROWTHER S A, MOHAPATRA R K, TURNER G, et al. Characteristics and applications of relax, an ultrasensitive resonance ionization mass spectrometer for xenon[J]. Journal of Analytical Atomic Spectrometry, 2008, 23(7): 938-947.
    [11]
    SAVINA M R, JELLIN M J, TRIPA C E, et al.Analyzing individual presolar grains with CHARISMA[J]. Geochimica et Cosmochimica Acta, 2003, 67(17): 3215-3225.
    [12]
    WUNDERLICH R K, WASSERBURG G J, HUTCHEON I D, et al. Laser-induced isotopic effects in Titanium resonance ionization[J]. Anal Chem, 1993, 65: 1411-1418.
    [13]
    BLAISE J. Term analysis of the spectrum of neutral Plutonium, Pu I[J]. Optical Society of America, 1986, 3(3): 403-518.
    [14]
    RIMKE H, HERRMANN G. Principle and analytical applications of resonance ionization mass spectrometry[J]. Mikrochim Acta[Wien], 1989, III: 223-230.
    [15]
    李志明. 激光共振电离质谱同位素定量分析技术[D]. 西安: 西北核技术研究所, 2005.
    [16]
    STRASHNOV I, BLAGBURN D J, GILMOUR J D. Hyperfine structure induced isotopic effects in krypton resonance ionization mass spectrometry[J]. Optics Communication, 2009, 282(17):3487-3492.
  • Related Articles

    [1]FENG Yuan-yuan, ZHOU Tao, TANG Yi-chuan, ZHANG Jian-ying, LI Peng-hui. Quantification of Trace Elements in Superalloy by Quadrupole ICP-MS Based on Oxygen Reaction Mode and Study on the Elimination Mechanism of Spectral Interference[J]. Journal of Chinese Mass Spectrometry Society, 2024, 45(4): 531-539. DOI: 10.7538/zpxb.2024.0001
    [2]ZHANG Feng, MU Guo-dong, LIU Tong. Research Progress of Highly Selective Solid Substrate Electrospray Ionization Sources for Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2022, 43(4): 405-417. DOI: 10.7538/zpxb.2021.0194
    [3]HAN Juan, LIU Han-bin, JIN Gui-shan, ZHANG Jian-feng, LI Jun-jie, ZHANG Jia, SHI Xiao. Carbon and Oxygen Isotope Analysis of Calcite and Dolomite Mixtures Using Selective Acid Extraction by Gasbench-IRMS[J]. Journal of Chinese Mass Spectrometry Society, 2022, 43(2): 210-219. DOI: 10.7538/zpxb.2021.0071
    [4]JIA He-yuan, YAO Bo, CHEN Shi-lyu, LU Shi-fang, CAO Jie. Free Radical-Mediated Selective C—C Bond Cleavage and Differentiation of Isomers in Peptide Derivatives[J]. Journal of Chinese Mass Spectrometry Society, 2022, 43(1): 44-55. DOI: 10.7538/zpxb.2021.0019
    [5]YING Zhe-cong, XIONG Chao-jie, ZHANG Li-hua, LI Hui-bo, HAO Xiao-juan. Application of Total Evaporation Technique Using Thermal Ionization Mass Spectrometry in Measurement of Plutonium Isotopes#br#[J]. Journal of Chinese Mass Spectrometry Society, 2021, 42(3): 305-313. DOI: 10.7538/zpxb.2020.0053
    [6]LIN Zheng, MENG Yi-fan, WANG Tong-tong, HANG Wei. Laser Postionization Technique for Enhancing Signals of High Ionization Potential Elements[J]. Journal of Chinese Mass Spectrometry Society, 2020, 41(2): 175-180. DOI: 10.7538/zpxb.2019.0165
    [7]LIU Chang-jing, MEI Hong-cheng, ZHU Jun, HE Ya, HE Xin-long, JIANG Jing, QUAN Yang-ke, WANG Gui-qiang. Selection of Reference Materials for Hydrogen Stable Isotope Detection of Human Hair[J]. Journal of Chinese Mass Spectrometry Society, 2019, 40(4): 342-348. DOI: 10.7538/zpxb.2018.0144
    [8]LI Bing, LIANG Bang-hong, SU Dong-ping, ZHANG Jing-song, CHEN Yun-ming, SUN Peng, LI Shun-tao, LI Xin-zheng, YUAN Ping. Determination of 239Pu and 240Pu for Radioactive Liquid Waste by High Resolution Inductively Coupled Plasma Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2018, 39(5): 593-598. DOI: 10.7538/zpxb.2017.0180
    [9]YAN Lai-lai, LIU Li-ping, NIU Gang, WENG Zhi-hui, WANG Jing-yu. Correlation Research between Lead and Other Elements in the Blood of Lead-Exposed Rats[J]. Journal of Chinese Mass Spectrometry Society, 2009, 30(5): 295-299.
    [10]YU Shi-dan. Determination of Organochlorine Pesticides by SPE Disk Extraction-GC/MS with Selected Ion Monitoring Mode[J]. Journal of Chinese Mass Spectrometry Society, 2006, 27(4): 209-214.

Catalog

    Article views (933) PDF downloads (1110) Cited by()

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return