YUAN Hong-zhao, ZHANG Li-ping, GENG Mei-mei, XU Li-wei, LI Chun-yong, SONG Li-fang, WANG Jiu-rong. Comparison of Methods for Hydrogen and Oxygen Isotopes Analysis of Water Samples by Flash HT and GasBench II-IRMS System[J]. Journal of Chinese Mass Spectrometry Society, 2013, 34(6): 347-352. DOI: 10.7538/zpxb.2013.34.06.0347
Citation: YUAN Hong-zhao, ZHANG Li-ping, GENG Mei-mei, XU Li-wei, LI Chun-yong, SONG Li-fang, WANG Jiu-rong. Comparison of Methods for Hydrogen and Oxygen Isotopes Analysis of Water Samples by Flash HT and GasBench II-IRMS System[J]. Journal of Chinese Mass Spectrometry Society, 2013, 34(6): 347-352. DOI: 10.7538/zpxb.2013.34.06.0347

Comparison of Methods for Hydrogen and Oxygen Isotopes Analysis of Water Samples by Flash HT and GasBench II-IRMS System

More Information
  • In order to establish a more accurately and rapid method for hydrogen and oxygen isotopes analysis of water sample, two commercially available equipments of elemental analyzer isotope ratio mass spectrometer (Flash HT-IRMS) and sample preparation device-stable isotope mass spectrometer (GasBench II-IRMS) with the reduction of H2O samples by reaction with glassy carbon at high temperatures and aqua equilibration techniques, respectively, were used for the analysis of hydrogen and oxygen isotopes of water samples from different sources. The stable isotope measurements results show that both of analytical systems are able to provide good analytical data in accuracy with precision of superior to 0.5‰ (δD) and 0.1‰ (δ18O), respectively. The comparison of the analytical data from the two analytical systems indicates that Flash HT-IRMS continuous flow method provides better hydrogen and oxygen isotope composition data in reproducibility and accuracy than the data from the other method. In addition, Flash HT-IRMS analysis system can on line simultaneous determination of δD and δ18O in water samples with short analysis time and sample consumption, only 0.1 μL. Therefore, for the analysis of hydrogen and oxygen isotopes in micro-amounts of water samples, the Flash HT-IRMS analysis system has more advantages.
  • [1]
    石辉, 刘世荣, 赵晓广. 稳定性氢氧同位素在水分循环中的应用[J]. 水土保持学报, 2003, 17(2): 163-166.SHI Hui, LIU Shirong, ZHAO Xiaoguang. Application of stable hydrogen and oxygen isotope in water circulation [J]. Journal of Soil Water Conservation, 2003, 17(2): 163-166(in Chinese).
    [2]
    白志鹏, 张利文, 朱坦, 等. 稳定同位素在环境科学研究中的应用进展[J]. 同位素, 2007, 20(1): 57-64.BAI Zhipeng, ZHANG Liwen, ZHU Tan, et al. The status of applying stable isotope in the studies of environmental science[J]. Journal of Isotopes, 2007, 20(1): 57-64(in Chinese).
    [3]
    AGGARWAL P P, GAT J R, FROEHLICH K F O. Isotopes in the water cycle: Past, present and future of a developing science[M]. Dordrecht: Springer, 2007: 381.
    [4]
    谭忠成, 陆宝宏, 汪集旸, 等. 同位素水文学研究综述[J]. 河海大学学报:自然科学版, 2009,37(1):16-22.TAN Zhongcheng, LU Baohong,WANG Jiyang, et al. Isotope hydrology: Progress and prospects[J]. Journal of Hohai University (Natural Sciences), 2009,37(1):16-22(in Chinese).
    [5]
    EILER J M, CLOG M, MAGYAR P, et al. A high-resolution gas-source isotope ratio mass spectrometer[J]. International Journal of Mass Spectrometry, 2013, 335: 45-56.
    [6]
    FLENKER U. Isotope ratio mass spectrometry-history and terminology in brief[J]. Drug Testing and Analysis, 2012, 4(12): 893-896.
    [7]
    陶成, 张美珍, 杨华敏, 等. GasBench-IRMS水平衡氢同位素分析方法研究及应用[J]. 质谱学报, 2006, 27(4): 215-219.TAO Cheng, ZHANG Meizhen, YANG Huamin, et al. The research and application of GasBench-IRMS H2/H2O (l) equlibrium H/D isotope analysis[J]. Journal of Chinese Mass Spectrometry Society, 2006, 27(4): 215-219(in Chinese).
    [8]
    孙青, 王晓华, 石丽明. GasBench II-IRMS水平衡氢氧同位素分析方法研究[J]. 岩矿测试, 2009, 28(1): 1-4.SUN Qing, WANG Xiaohua, SHI Liming, et al. Determination of hydrogen and oxygen isotope ratios in water samples by GasBench Ⅱ-IRMS using aqua equilibration technique[J]. Rock and Mineral Analysis, 2009, 28(1): 1-4(in Chinese).
    [9]
    UEMURA R, MATSUIY Y, MOTOYAMA H, et al. Deuterium and oxygen-18 determination of microliter quantities of a water sample using an automated equilibrator [J]. Rapid Communications in Mass Spectrometry, 2007, 21: 1783-1790.
    [10]
    LU F H. On line high-precision D2H and D18O analysis in water by pyrolysis[J]. Rapid Communications in Mass Spectrometry, 2009, 23: 3144-3150.
    [11]
    张琳, 陈宗宇, 刘福亮, 等. 水中氢氧同位素不同分析方法的对比[J]. 岩矿测试, 2011, 30(2): 160-163.ZHANG Lin, CHEN Zongyu, LIU Fuliang, et al. Study on methods for hydrogen and oxygen isotope analysis of water samples[J]. Rock and Mineral Analysis, 2011, 30(2): 160-163(in Chinese).
    [12]
    SHOUAKAR-STASH O, DRIMMIE R, MORRISON J, et al. On-line D/H analysis for water, natural gas, and organic solvents by manganese reduction[J]. Analytical Chemistry, 2000, 2: 2664-2666.
    [13]
    SIMON D K, KARL D H, PAUL B. Deuterium/hydrogen isotope ratio measurement of water and organic samples by continuous-flow isotope ratio mass spectrometry using chromium as the reducing agent in an elemental analyzer[J]. Rapid Communications in Mass Spectrometry, 2001, 15(15): 1283-1286.
    [14]
    BARKAN E, LUZ B. High precision measurements of 17O/16O and 18O/16O ratios in H2O[J]. Rapid Communications in Mass Spectrometry, 2005, 19(24): 3737-3742.
    [15]
    MCLLVIN M R, CASCIOTTI K L. Method for the analysis of D18O in water[J]. Analytical Chemistry, 2006, 78: 2377-2381.
    [16]
    刘运德, 甘义群, 余婷婷, 等. 微量水氢氧同位素在线同时测试技术热转换元素分析同位素比质谱法[J]. 岩矿测试, 2010, 29(6): 643-647.LIU Yunde, GAN Yiqun, YU Tingting, et al. Online simultaneous determination of δD and δ18O in Micro-liter water samples by thermal conversion/elemental analysis-isotope ratio mass spectrometry[J]. Rock and Mineral Analysis, 2010, 29(6): 643-647(in Chinese).

Catalog

    Article views (970) PDF downloads (1160) Cited by()

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return