LI Gang, LI De-tian, CHENG Yong-jun, PEI Xiao-qiang, WANG Yong-jun, ZHANG Hu-zhong, DONG Meng, SUN Jian, ZHANG Qi. Effect of Voltage Instability on Motion Characteristics of Ions in Ion Trap[J]. Journal of Chinese Mass Spectrometry Society, 2018, 39(1): 1-9. DOI: 10.7538/zpxb.2017.0034
Citation: LI Gang, LI De-tian, CHENG Yong-jun, PEI Xiao-qiang, WANG Yong-jun, ZHANG Hu-zhong, DONG Meng, SUN Jian, ZHANG Qi. Effect of Voltage Instability on Motion Characteristics of Ions in Ion Trap[J]. Journal of Chinese Mass Spectrometry Society, 2018, 39(1): 1-9. DOI: 10.7538/zpxb.2017.0034

Effect of Voltage Instability on Motion Characteristics of Ions in Ion Trap

More Information
  • Resonant excitation is a widely used ion ejection technique in ion trap mass spectrum. Combined with mass selective instability scan mode, it can remarkably improve mass resolution and scanning range. The performances of ion trap mass spectrometers, such as mass resolution, sensitivity etc., depend on the motion characteristics of ions, which is vulnerable to both initial conditions and external field parameters. Small changes in voltage can lead to observably oscillation in the motion of ions confined in ion traps. To explore this effect, a theoretical model was established based on the pseudo potential theory. Taking the burrs of the additional auxiliary AC voltage and the DC voltage applied on end cap electrodes as examples, the influence of the stability of voltage on the performance of ion trap was revealed using this model. The results showed that if the auxiliary AC voltage existed burrs, sudden changes in ion velocity and motion amplitude would occur, leading to disruption of the cooling state of ions in ion trap, some of which would hit the trap wall in the excitation phase and could not be detected, thereby reducing the sensitivity of the ion trap mass spectrometer. In addition, the model showed that the secular frequency could be determined through the measurement of the response to the action of voltage burr. By applying a pulse voltage on the ion trap electrode, measuring the ion response to the excitation, and implementing Fourier transformation to the time-zone response, the frequency corresponding to the peak value of the frequency-zone curve was the secular frequency. If a DC voltage was applied on the end cap electrodes, the equilibrium position of the ion motion would be significantly shifted, the amplitude of which was considerable compared with the small size of the ion trap. It could be expected that more than 50% of the ions would eject from the hatch of the end cap electrode on the offset side. Therefore, placing the ion detector outside this electrode can greatly improve the detection efficiency, and hence improve the sensitivity of the ion trap mass spectrometer. The theoretical model established in this work can reasonably explain experimental phenomena, proving its effectiveness. This work can provide theoretical reference for the optimal design of ion trap mass spectrometer. During the actual design and implementation of the ion trap mass spectrometer, the RF signal is more prone to burr than the additional auxiliary AC voltage. In the future, it will have important practical significance to study the influence of the stability of RF signal on the ion motion characteristics in ion trap.
  • [1]
    LOURIS J N, COOKS R G, SYKA J E P, et al. Instrumentation, applications, and energy deposition in quadrupole ion-trap tandem mass spectrometry[J]. Analytical Chemistry, 1987, 59(13): 1677-1685.
    [2]
    STAFFORD G C, KELLEY P E, SYKA J E P, et al. Recent improvements in and analytical applications of advanced ion trap technology[J]. International Journal of Mass Spectrometry and Ion Processes, 1984, 60(1): 85-98.
    [3]
    范荣荣,薛兵,孙露露,等. 高压模块性能对数字离子阱质谱仪性能的影响[J]. 质谱学报,2015,36(2):97-103.FAN Rongrong, XUE Bing, SUN Lulu, et al. Effect of high voltage module performance on digital ion trap mass spectrometer[J]. Journal of Chinese Mass Spectrometry Society, 2015, 36(2): 97-103(in Chinese).
    [4]
    党乾坤,杨凯,王强,等. 利用直流偏置电压提高矩形离子阱质谱的灵敏度[J]. 质谱学报,2014,35(3):210-215.DANG Qiankun, YANG Kai, WANG Qiang, et al. Application of a DC offset in a rectilinear ion trap to improve sensitivity[J]. Journal of Chinese Mass Spectrometry Society, 2014, 35(3): 210-215(in Chinese).
    [5]
    WUERKER R F, SHELTON H, LANGMUIR R V. Electrodynamic containment of charged particles[J]. Journal of Applied Physics, 1959, 30(3): 342-349.
    [6]
    DEHMELT H G. Advances in atomic and molecular physics[J]. DR Bates, Academic, New-York, 1967.
    [7]
    ZHOU X, ZHU Z, XIONG C, et al. Characteristics of stability boundary and frequency in nonlinear ion trap mass spectrometer[J]. Journal of the American Society for Mass Spectrometry, 2010, 21(9): 1588-1595.
    [8]
    WHITTEN W B, REILLY P T A, RAMSEY J M. High-pressure ion trap mass spectrometry[J]. Rapid Communications in Mass Spectrometry, 2004, 18(15): 1749-1752.
    [9]
    GOERINGER D E, WHITTEN W B, RAMSEY J M, et al. Theory of high-resolution mass spectrometry achieved via resonance ejection in the quadrupole ion trap[J]. Analytical Chemistry, 1992, 64(13): 1434-1439.
    [10]
    ARNOLD N S, HARS G, MEUZELAAR H L C. Extended theoretical considerations for mass resolution in the resonance ejection mode of quadrupole ion trap mass spectrometry[J]. Journal of the American Society for Mass Spectrometry, 1994, 5(7): 676-688.
    [11]
    MAKAROV A A. Resonance ejection from the Paul trap: a theoretical treatment incorporating a weak octapole field[J]. Analytical Chemistry, 1996, 68(23): 4257-4263.
    [12]
    ZHAO X Z, GRANOT O, DOUGLAS D J. Quadrupole excitation of ions in linear quadrupole ion traps with added octopole fields[J]. Journal of the American Society for Mass Spectrometry, 2008, 19(4): 510-519.
    [13]
    XU W, SONG Q, SMITH S A, et al. Ion trap mass analysis at high pressure: a theoretical view[J]. Journal of the American Society for Mass Spectrometry, 2009, 20(11): 2144-2153.
    [14]
    RAJANBABU N, CHATTERJEE A, MENON A G. Motional coherence during resonance ejection of ions from Paul traps[J]. International Journal of Mass Spectrometry, 2007, 261(2): 159-169.
    [15]
    XU W, CHAPPELL W J, OUYANG Z. Modeling of ion transient response to dipolar AC excitation in a quadrupole ion trap[J]. International Journal of Mass Spectrometry, 2011, 308(1): 49-55.
    [16]
    MARCH R E, TODD J F. Quadrupole ion trap mass spectrometry[M]. John Wiley & Sons, 2005.