JIANG Pei-he, ZHAO Zhan-feng. Simulation on a Novel Method for Low Vacuum Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2020, 41(6): 588-594. DOI: 10.7538/zpxb.2020.0046
Citation: JIANG Pei-he, ZHAO Zhan-feng. Simulation on a Novel Method for Low Vacuum Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2020, 41(6): 588-594. DOI: 10.7538/zpxb.2020.0046

Simulation on a Novel Method for Low Vacuum Mass Spectrometry

More Information
  • As an advanced product in scientific instruments, mass spectrometer plays an important role in many areas. Recently, a lot of efforts have been put on the miniature ion trap MS system development. However, due to the restraints of vacuum systems, it is difficult to reduce the size of a miniature mass spectrometer furthermore. Performing mass analysis in a low-vacuum environment can avoid the use of a bulky turbo pump, which will significantly reduce the size, weight, cost and power of the instrument. Moreover, this technology can be used to develop next generation hand-portable instruments and bring mass spectrometry to new applications. Quadrupole mass filter is one of the most mature and widely used mass analysers. In general, when using the quadrupole mass filter, the pressure in the chamber needs to be maintained on the order of 10-5 Pa. As background pressure increasing, the ions collide with the background gas molecules more frequently. The energy of the ion's axial motion decreases gradually and finally falls to zero. Thus, mass analysis failed since the ions are unable to reach the detector located at the end of the analyser. To address this issue, a mass analysis method was proposed in this study where the drift gas flow was used to provide the ejection energy for ions. During mass analysis, the sample was ionised in the ion source. Then, the ions shot into the mass analyser and travelled under the combined action of the quadrupole field and the drift gas flow. The drift gas flow provided the kinetic energy for axial ion motion, and the quadrupole field selected ions depending on the mass-to-charge ratio in the cross-sectional direction. Thus, although in low-vacuum environment, ions in stable motion could pass through the mass analyser and hit the ion detector, where spectral information was generated. The simulation platform was built in COMSOL Multiphysics. The simulation results showed that the obtained spectral peak was superior to those without drift gas flow in the intensity and the width of the spectral peak. The proposed method is suitable when a high precision is not required. Besides, as components for low vacuum mass spectrometry, a novel quadrupole mass filter and a novel ion detector were used. For the novel QMF, rectangular electrodes were employed instead of traditional hyperboloid or cylindrical electrodes. To eliminate the turbulence near the ion detector when the drift gas flow through the detector, a novel ion detector with a double layer structure was constructed. The research results lay a theoretical foundation for low vacuum mass spectrometer, and it will promote the development of portability mass spectrometers. In the following work, a prototype will be built to verify the simulation results.
  • [1]
    GAO L, SONG Q, PATTERSON G E, COOKS R G, OUYANG Z. Handheld rectilinear ion trap mass spectrometer[J]. Analytical Chemistry, 2006, 78(17): 5 994-6002.
    [2]
    GAO L, SUGIARTO A, HARPER J D, COOKS R G, OUYANG Z. Design and characterization of a multisource hand-held tandem mass spectrometer[J]. Analytical Chemistry, 2008, 80(19): 7198-7205.
    [3]
    LI L, CHEN T, REN Y, HENDRICKS P I, COOKS R G, OUYANG Z. Miniature mass spectrometer for clinical and other applications-introduction and characterization[J]. Analytical Chemistry, 2014, 86(6): 2909-2916.
    [4]
    HE M, XUE Z, ZHANG Y, HUANG Z, FANG X, QU F, OUYANG Z, XU W. Development and characterizations of a miniature capillary electrophoresis mass spectrometry system[J]. Analytical Chemistry, 2015, 87(4): 2236-2241.
    [5]
    ZHAI Y, FENG Y, WEI Y, WANG Y, XU W. Development of a miniature mass spectrometer with continuous atmospheric pressure interface[J]. Analyst, 2015, 140(10): 3406-3414.
    [6]
    BRODBELT J S. Effects of collisional cooling on ion[J]. Practical Aspects of Ion Trap Mass Spectrometry, 1995(1): 209-217.
    [7]
    闫岩,范荣荣,薛兵,孙露露,董俊国,高伟,黄正旭,程平,周振. 缓冲气体冷却对便携式数字线性离子阱质谱仪性能影响的研究[J]. 现代科学仪器,2014(6):55-59.YAN Yan, FAN Rongrong, XUE Bing, SUN Lulu, DONG Junguo, GAO Wei, HUANG Zhengxu, CHENG Ping, ZHOU Zhen. Buffer gas cooling effect on portable digital linear ion trap mass spectrometer[J]. Modern Scientific Instruments, 2014(6): 55-59(in Chinese).
    [8]
    XU W, SONG Q, SMITH S A, CHAPPELL W J, OUYANG Z. Ion trap mass analysis at high pressure: a theoretical view[J]. Journal of the American Society for Mass Spectrometry, 2009, 20(11): 2144-2153.
    [9]
    SONG Q, XU W, SMITH S A, GAO L, CHAPPELL W J, COOKS R G, OUYANG Z. Ion trap mass analysis at high pressure: an experimental characterization[J]. Journal of Mass Spectrometry, 2010, 45(1): 26-34.
    [10]
    EUGENE A, VINITSKY E D B K. Particle dynamics in damped nonlinear quadrupole ion traps[J]. American Journal of Physics, 2015, 83(4): 313-319.
    [11]
    ZIAEIAN I, NOSHAD H. Theoretical study of the effect of damping force on higher stability regions in a Paul trap[J]. International Journal of Mass Spectrometry, 2009, 289(1): 1-5.
    [12]
    WHITTEN W B, REILLY P T A, RAMSEY J M. High-pressure ion trap mass spectrometry[J]. Rapid Communications in Mass Spectrometry, 2004, 18(15): 1749-1752.
    [13]
    NASSE M, FOOT C. Influence of background pressure on the stability region of a Paul trap[J]. European Journal of Physics, 2001, 22(6): 563-573.
    [14]
    姜佩贺,赵占锋,陈焕文,周志权. 低真空度下四极杆质量分析器稳定区的数值计算[J]. 哈尔滨工业大学学报,2017,49(5):31-35.JIANG Peihe, ZHAO Zhanfeng, CHEN Huanwen, ZHOU Zhiquan. Numerical method for stability region in quadrupole at low vacuum[J]. Journal of Harbin Institute of Technology, 2017, 49(5): 31-35(in Chinese).
    [15]
    JIANG P, ZHOU Z, WU Z, ZHAN Z. Low-vacuum cylindrical ion trap mass spectrometry[J]. Instrumentation Science and Technology, 2018, 46(6): 614-627.
    [16]
    姜佩贺,周志权,陈焕文,赵占锋. 低真空环境下四极杆质量分析器灵敏度仿真研究[J]. 质谱学报,2018,39(1):20-27.JIANG Peihe, ZHOU Zhiquan, CHEN Huanwen, ZHAO Zhanfeng. Simulation of sensitivity in quadrupole mass spectrometer at low vacuum[J]. Journal of Chinese Mass Spectrometry Society, 2018, 39(1): 20-27(in Chinese).
    [17]
    JIANG P, ZHOU Z, WU S, ZHAO Z. A novel quadrupole mass filter based on rectangular electrodes[C]. International Conference on Instrumentation and Measurement, Computer, Communication and Control(IMCCC 2018), China: Harbin, 2018: 1524-1527.
    [18]
    TANG Y, XU Q, LI D, XU W. A mini mass spectrometer with a low noise Faraday detector[J]. Analyst, 2020, 145(11): 3892-3898.
    [19]
    JIANG P, ZHOU Z, ZHAO Z. A novel ion detector aiming at the homogeneous of drift gas[J]. International Journal for Ion Mobility Spectrometry, 2017, 20(3): 111-117.

Catalog

    Article views (1020) PDF downloads (607) Cited by()

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return