Effect of Amplitude-Scanning ac on the Performance of Ion Trap Mass Spectrometry

More Information
  • With the wide application of ion trap mass spectrometry, the theoretical simulation and experimental verification of resonant ejection play an important role in improving the performance of ion trap mass spectrometry. There are two common ways to set the (alternating current) (ac) resonant voltage of ion trap, one is to set a small constant value like 1 V0-p, and the other is to set an amplitude-scanning range. However, few people has studied the influence of the amplitude-scanning setting of ac on the analytical performance of ion trap mass spectrometry, and no experimental result has been studied by theoretical simulation neither. In this work, based on a homemade miniaturized ion trap mass spectrometry platform with continuous atmospheric pressure interface, the effect of amplitude-scanning resonance on the performance of ion trap mass spectrometry was studied by theoretical simulation and experimental study using a nano electrospray ionization source, and reserpine was used as the research object. Firstly, the effects of constant ac resonant ejection and amplitude-scanning ac resonant ejection on the resolution and sensitivity of ion trap mass spectrometry were studied experimentally. The experimental results showed that amplitude-scanning resonant ac ejection could improve the resolution and sensitivity of ion trap mass spectrometry, SIMION showed that the amplitude-scanning ac resonant ejection could prevent the ions from being excited in advance compared with the constant ac resonant ejection subsequently. Finally, the application experiments of two kinds of ions, fentanyl and narcotine showed that the resolution of amplitude-scanning ac resonant ejection was more than 2 times higher than that of constant ac resonant ejection. In conclusion, the effect of amplitude-scanning ac on the performance of ion trap mass spectrometry was studied by theoretical simulation and experimental comparison, and the theoretical simulation method and related program of resonance excitation process of ion trap mass spectrometry were established, which laid a foundation for further study of more complex nonlinear resonance and other high-order motion in ion trap. At the same time, the conclusions also help to improve the analytical performance of ion trap mass spectrometry to a certain extent, and accelerate the instrument debugging process of ion trap mass spectrometry.
  • [1]
    PAUL W, OSBERGHAUS O, FISCHER E. Forschungsberichte des wirtchaft und verkehrministeriums nordrhein westfalen[J]. Westdeutscher Verlag, Kln and Opladen, 1958: 415.
    [2]
    DAWSON P H, WHETTEN N R. Ion storage in threedimensional, rotationally symmetric, quadrupole fields. I. Theoretical treatment[J]. Journal of Vacuum Science&Technology, 1968, 5: 1-10.
    [3]
    MARCH R E, TODD J F. Quadrupole ion trap mass spectrometry[M]. John Wiley & Sons: 2005: 165.
    [4]
    MARCH R E, TODD J F. Practical aspects of trapped ion mass spectrometry, volume Ⅳ: theory and instrumentation[M]. CRC Press, 2010.
    [5]
    HENDRICKS P, DUNCAN J, NOLL R J, OUYANG Z, COOKS R G. Performance of a low voltage ion trap[J]. International Journal of Mass Spectrometry, 2011, 305(1): 69-73.
    [6]
    TAYLOR N, AUSTIN D E. A simplified toroidal ion trap mass analyzer[J]. International Journal of Mass Spectrometry, 2012, 321: 25-32.
    [7]
    SCHWARTZ J C, SENKO M W, SYKA J E. A twodimensional quadrupole ion trap mass spectrometer[J]. Journal of the American Society for Mass Spectrometry, 2002, 13(6): 659-669.
    [8]
    DANG Q, XU F, XIE X, XU C, DING C F. Enhancement of ion activation and collisioninduced dissociation by simultaneous dipolar excitation of ions in X- and Y-directions in a linear ion trap[J]. Analytical Chemistry, 2015, 87(11): 5561-5568.
    [9]
    LI X, JIANG G, LUO C, XU F, DING C F. Ion trap array mass analyzer: structure and performance[J]. Analytical Chemistry, 2009, 81(12): 4840-4846.
    [10]
    JONSCHER K R, YATES J R. The quadrupole ion trap mass spectrometera small solution to a big challenge[J]. Analytical Biochemistry, 1997, 244(1): 1-15.
    [11]
    WANG Y, DING C F, XU F, KONENKOV N V, DING H. Linear ion trap mass selectivity with impulse power supply and sinusoidal dipolar excitation[J]. Rapid Communications in Mass Spectrometry, 2016, 30: 2664-2670.
    [12]
    JIANG T, XU Q, ZHANG H, LI D, XU W. Improving the performances of a “brick mass spectrometer” by quadrupole enhanced dipolar resonance ejection from the linear ion trap[J]. Analytical Chemistry, 2018, 90(19): 11671-11679.
    [13]
    SNYDER D T, PULLIAM C J, WILEY J S, DUNCAN J, COOKS R G. Experimental characterization of secular frequency scanning in ion trap mass spectrometers[J]. Journal of The American Society for Mass Spectrometry, 2016, 27(7): 1243-1255.
    [14]
    SCHWARTZ J C, SYKA J E, JARDINE I. High resolution on a quadrupole ion trap mass spectrometer[J]. Journal of the American Society for Mass Spectrometry, 1991, 2(3): 198-204.
    [15]
    MOXOM J, REILLY P T, WHITTEN W B, RAMSEY J M. Double resonance ejection in a micro ion trap mass spectrometer[J]. Rapid Communications in Mass Spectrometry, 2002, 16(8): 755-760.
    [16]
    WILLIAMS J D, COX K A, COOKS R G, McLUCKEY S A, HART K J, GOERINGER D E. Resonance ejection ion trap mass spectrometry and nonlinear field contributions: the effect of scan direction on mass resolution[J]. Analytical Chemistry, 1994, 66(5): 725-729.
    [17]
    XU F, WANG W, JIN L, QIAN B, DING C F. Measurement of the effective electric field radius on digital ion trap spectrometer[J]. Analyst, 2021, 146: 3810-3817.
    [18]
    ZHANG Z Y, YUAN G Z, HE Y, QIAN J, ZHANG S G, YAO R J, DING C F, LI X X. Research of unidirectional ion ejection in printed-circuit-board ion trap[J]. Chinese Journal of Analytical Chemistry, 2017, 45(2): 297-302.
  • Cited by

    Periodical cited type(1)

    1. 裴泓明,刘梅英,徐赫奕,张谛,谢洁,屈子裕,黄泽建,蒋长元,蒙颖艳,戴新华,方向,王庆辉,江游. 离子阱质谱仪质量分段校正算法研究. 质谱学报. 2023(05): 676-684 . 本站查看

    Other cited types(1)

Catalog

    Article views (418) PDF downloads (279) Cited by(2)

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return