LIU Xue-mei, TANG Lei, LONG Kai-ming, FU Zhong-hua, SHAO Xue-peng. Calibration Technique of the Fractionation Effect in High Precision Isotopic Ratio Analysis for Trace Lead[J]. Journal of Chinese Mass Spectrometry Society, 2013, 34(5): 282-287. DOI: 10.7538/zpxb.2013.34.05.0282
Citation: LIU Xue-mei, TANG Lei, LONG Kai-ming, FU Zhong-hua, SHAO Xue-peng. Calibration Technique of the Fractionation Effect in High Precision Isotopic Ratio Analysis for Trace Lead[J]. Journal of Chinese Mass Spectrometry Society, 2013, 34(5): 282-287. DOI: 10.7538/zpxb.2013.34.05.0282

Calibration Technique of the Fractionation Effect in High Precision Isotopic Ratio Analysis for Trace Lead

  • Fractionation effects were corrected by empirical (K correction) and double-dilution methods in isotopic measurements of 1—50 ng lead standard sample. Results showed that the relative deviations of 206Pb/204Pb, 207Pb/204Pb, 208Pb/204Pb could be reduced from 0.1%—0.6% to 0.02%—0.38% for Pb sample more than 10 ng using empirical methods, and using double-dilution methods it could be reduced from 0.08%—0.45% to 0.001%—0.14%. But no obvious effect were detected for Pb sample less than 10 ng of two methods, which implied other facetors that affceted the precision besides fractionation effects.