SHI Ya-jing, LIU Hong, WANG Tong, XU Jin-yong, XIANG Mei, REN Zi-wei. Design of Double Sampling Loops System of Inductively Coupled Plasma Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2019, 40(3): 253-258. DOI: 10.7538/zpxb.2018.0097
Citation: SHI Ya-jing, LIU Hong, WANG Tong, XU Jin-yong, XIANG Mei, REN Zi-wei. Design of Double Sampling Loops System of Inductively Coupled Plasma Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2019, 40(3): 253-258. DOI: 10.7538/zpxb.2018.0097

Design of Double Sampling Loops System of Inductively Coupled Plasma Mass Spectrometry

More Information
  • Inductively coupled plasma mass spectrometry (ICP-MS) is usually coupled with solution pneumatic nebulizer sampling system, whose detection efficiency is low when measured in series mode. Based on the double-layer six-way multi-way valve, a double sample ring sampling system was designed, which could improve the detection efficiency by changing the connection mode of the sampling system. This system connected with pneumatic nebulizer, one loop was loading sample, the other loop was detecting. The detection time was saved by two-way parallel mode. Determination of different numbers of elements needed different length loop. The correlation coefficients of calibration curves of Sc, Ti, V, Cr, Mn, As, Co, Ni, Cu, Pb, Zn, Cd, Li, Sr, Ba, Co, U, and Th are greater than 0.999, and the precisions (n=5) are less than 5%. 1, 6, 12, 18, and 32 elements were determinted, their detection time was saved 45%, 35%, 27%, 21%, 16%, respectively. The number of elements is fewer, the improvement of the efficiency is higher. Compared with standard method, the limits of detection and sensitivity of this system have no significant difference. The double sampling loops system has no significant memory effect. The two-layer six-way multi-way valve switched manually and owned automation prospect. This system is high efficient, fast and simple, easy to realize automation, and more possibilities to use with other instruments.
  • [1]
    郭冬发,李金英,李伯平,谢胜凯,谭靖,张彦辉,刘瑞萍. 电感耦合等离子体质谱分析方法的重要进展(2005~2016年)[J]. 质谱学报,2017,38(5):599-610.GUO Dongfa, LI Jinying, LI Boping, XIE Shengkai, TAN Jing, ZHANG Yanhui, LIU Ruiping. Important progress in inductively coupled plasma mass spectrometry (2005-2016)[J]. Journal of Chinese Mass Spectrometry Society, 2017, 38(5): 599-610(in Chinese).
    [2]
    李金英,郭冬发,姚继军,曹淑琴. 电感耦合等离子体质谱(ICP-MS)新进展[J]. 质谱学报,2002,23(3):164-169.LI Jinying, GUO Dongfa, YAO Jijun, CAO Shuqin. New progress in inductively coupled plasma mass spectrometry (ICP-MS)[J]. Journal of Chinese Mass Spectrometry Society, 2002, 23(3): 164-169(in Chinese).
    [3]
    丁兰,金海燕,刘淼,梁枫,郇延富,曹彦波,张寒琦,金钦汉. 用低功率微波热雾化器与十字交叉气动雾化器作为ICP-AES进样装置的比较[J]. 高等学校化学学报,2000,21(11):1661-1663.DING Lan, JIN Haiyan, LIU Miao, LIANG Feng, XUN Yanfu, CAO Yanbo, ZHANG Hanqi, JIN Qinhan. Comparison of low power microwave thermal atomizer and cross-pneumatic atomizer as ICP-AES sampling device[J]. Chemical Journal of Chinese Universities, 2000, 21(11): 1661-1663(in Chinese).
    [4]
    白岗,孔祥丽,霍阳. 一种雾化去溶捕集样品引入系统及方法:CN,CN 103512944A[P]. 2014.
    [5]
    李冰,杨红霞. 电感耦合等离子体质谱仪原理和应用[M]. 北京:地质出版社,2005.
    [6]
    Tsuyoshi Morikawa, Tsutomu Okoba. Autosampler: USA. US09638674B2[P]. 2017-05-02.

Catalog

    Article views (880) PDF downloads (632) Cited by()

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return