LIU Fang, ZHU Hong-li, TAN De-can, LIU Yu-fei, KANG Jin-ting, ZHU Jian-ming, WANG Gui-qin, ZHANG Zhao-feng. Optimization of Calcium Double-Spike on Measuring Calcium Isotopic Compositions by Thermal Ionization Mass Spectrometer (TIMS)[J]. Journal of Chinese Mass Spectrometry Society, 2016, 37(4): 310-318. DOI: 10.7538/zpxb.youxian.2016.0031
Citation: LIU Fang, ZHU Hong-li, TAN De-can, LIU Yu-fei, KANG Jin-ting, ZHU Jian-ming, WANG Gui-qin, ZHANG Zhao-feng. Optimization of Calcium Double-Spike on Measuring Calcium Isotopic Compositions by Thermal Ionization Mass Spectrometer (TIMS)[J]. Journal of Chinese Mass Spectrometry Society, 2016, 37(4): 310-318. DOI: 10.7538/zpxb.youxian.2016.0031

Optimization of Calcium Double-Spike on Measuring Calcium Isotopic Compositions by Thermal Ionization Mass Spectrometer (TIMS)

  • Detailed investigations on how to choose the optimal double-spike and the optimal mixture ratios of the two single spikes in the double-spike, as well as on how to determinate the optimal sample/spike ratio range during measurement of calcium isotopic compositions by thermal ionization mass spectrometer (TIMS), were carefully carried out in both theory and practices. Compared to the 42Ca-48Ca and 43Ca-48Ca double-spikes, which are used widely, 42Ca-43Ca double-spike has its distinctive advantages. Firstly, although the error on the fractionation factor α is relatively bigger when using 42Ca-43Ca double-spike, but it only fluctuates a little at a much wide range of sample/spike ratio in spiked samples than using the 42Ca-48Ca or 43Ca-48Ca. Secondly, the average mass of the two isotopes in 42Ca-43Ca double-spike and target 44Ca/40Ca ratio, only differs by 0.5 mass unit from each other, thus errors in the corrected ratio resulted from fractionation correction will be minimized, especially when a lower amount of sample was loaded and/or the large fractionation on the instrument occurred during mass determination. Finally, 42Ca and 43Ca can be simultaneously collected by TIMS without significant ion optic effect, which can enhance the data quality collected. In summary, 42Ca-43Ca double-spike should be the optimal one to be used in the double-spike technique for precisely measuring calcium isotopic compositions by TIMS.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return