WANG Chen, ZHAO Yong-gang, ZHANG Ji-long, JIANG Xiao-yan, CHANG Zhi-yuan, ZHU Liu-chao. Determination of Uranium in Soil Samples by Flow Injection-ICP-MS[J]. Journal of Chinese Mass Spectrometry Society, 2010, 31(1): 34-38.
Citation: WANG Chen, ZHAO Yong-gang, ZHANG Ji-long, JIANG Xiao-yan, CHANG Zhi-yuan, ZHU Liu-chao. Determination of Uranium in Soil Samples by Flow Injection-ICP-MS[J]. Journal of Chinese Mass Spectrometry Society, 2010, 31(1): 34-38.

Determination of Uranium in Soil Samples by Flow Injection-ICP-MS

  • A rapid on-line separation and preconcentration method for the determination of uranium in soil was developed by flow injection multi-collector inductively coupled plasma mass spectrometry(FI-MC-ICP-MS) using 233U as spike for the isotope dilution analysis. The on-line separation was based on uranium adsorption on UTEVA resin. The absorbed uranium was eluted with 0.2 mol•L-1 HCl and directly introduced into MC-ICP-MS for measuring uranium isotopes. The method was examined by determining the uranium in soil standard reference materials. The results are in good agreement with the certified values. The limits of detection of 238U and 235U are 5.4×10-9 g and 5.7×10-11g.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return