ZHU Ming-yan, CUI Jian-yong, LIU Yu-ang, GUO Dong-fa. Quantitative Analysis of Uranium in Environmental Samples by ID-TIMS[J]. Journal of Chinese Mass Spectrometry Society, 2010, 31(增刊): 83-86.
Citation: ZHU Ming-yan, CUI Jian-yong, LIU Yu-ang, GUO Dong-fa. Quantitative Analysis of Uranium in Environmental Samples by ID-TIMS[J]. Journal of Chinese Mass Spectrometry Society, 2010, 31(增刊): 83-86.

Quantitative Analysis of Uranium in Environmental Samples by ID-TIMS

  • ID-TIMS technique was used to measure the uranium content and uranium isotope ratios for environmental samples, in which 233U was used as spike. Compared with traditional methods in which 235U is used as spike, this method simplifies the analysis steps. Furthermore, the results of uranium isotope ratios calculated by ID-TIMS technique were well consistent with those measured directly by TIMS. For ng level uranium the precision of 235U/238U was less than 0.7%, that of 234U/238U less than 10%.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return