The Determination of Hg Content in Polypropylene by ID-ICP-MS
-
-
Abstract
The content of Hg in polypropylene sample was determined by isotope dilution ICP-MS. The sample was digested by high-pressure vessel with HNO3/H2O2 as the solvent. By measuring 200Hg/202Hg abundance ratio and adding quantity of the spike, the content of Hg was determined. Moreover, the uncertainty of results were also analyzed and evaluated. The method indicates good accuracy, small uncertainty and less Hg loss during the pretreatment process. It is suitable for the certification of reference materials in national metrology institute.
-
-