WANG Chen, ZHAO Yong-gang, WANG Tong-xing, ZHANG Ji-long, CHANG Zhi-yuan, LI Li-li, ZHU Liu-chao, CHEN Yan. Determination of Lead in Blood by Isotope Dilution Multicollector Inductively Coupled Plasma Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2008, 29(2): 84-87.
Citation: WANG Chen, ZHAO Yong-gang, WANG Tong-xing, ZHANG Ji-long, CHANG Zhi-yuan, LI Li-li, ZHU Liu-chao, CHEN Yan. Determination of Lead in Blood by Isotope Dilution Multicollector Inductively Coupled Plasma Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2008, 29(2): 84-87.

Determination of Lead in Blood by Isotope Dilution Multicollector Inductively Coupled Plasma Mass Spectrometry

More Information
  • Corresponding author:

    WANG Chen

  • Received Date: July 10, 2007
  • Revised Date: November 28, 2007
  • A method for determination of trace Pb in blood was established by isotope dilution multicollector inductively coupled plasma mass spectrometry (MC-ICP-MS). The average Pb analytical blank obtained in this laboratory is 5.6 ng. A Nist standard reference materials of (SRM) 955b level 1 and (SRM) 955b level 3 were used to evaluate the accuracy of Pb concentration in blood. The measured values are 40.2 and 207.8 μg•L-1 respecetively, in good agreement with their certified value. The established method was used for determination the Pb concentration of three blood samples, the uncertainty of measurement results were also analyzed and evaluated . The Pb concentration of three blood samples are (102.7±5.5), (181.7±4.0) and (304.1±3.9) μg•L-1.
  • [1]
    凌 霞,毛云中.石墨炉原子吸收光谱法测定血样中铅[J].中国职业医学,2005,32(6):43-44.
    [2]
    金米聪,王 立,陈晓红.微分电位溶出法快速测定全血中铅[J].理化检验:化学分册,2002,38(6):293-294.
    [3]
    叶 涛,赵志成,吕 莉.氢化物发生原子荧光法测定全血铅含量[J].中华检验医学杂志,2003,26(11):677-679.
    [4]
    冯俊华,余远碧,黎 超.示波极谱法测定全血和尿样中铅[J].理化检验:化学分册,1999,35(2):82-84.
    [5]
    MURPHY K E, PAULSEN P J. The determination of lead in blood using isotope dilution inductively coupled plasma mass spectrometry[J]. Fresenius J Anal Chem, 1995, 352(1/2): 203-208.
    [6]
    刘立译.量化分析测量不确定度指南[M] . 北京: 中国计量出版社,2003:113.
    [7]
    刘虎生,邵宏翔. 电感耦合等离子体质谱技术与应用[M]. 北京:化学工业出版社,2005:170-173.
    [8]
    赵墨田,曹永明,陈 刚,等. 无机质谱法[M]. 北京:化学工业出版社,2005:175189.
    [9]
    中国合格评定国家认可委员会.CNASGL06: 2006 化学分析中不确定度的评估指南[S]. 2006:142-147.
    [10]
    王雪莹. 标准溶液测量不确定度的评定[J].理化检验:化学分册,2004,40(7):392-394.
  • Related Articles

    [1]WANG Wei, XU Jiang, ZHAI Li-hua, SHEN Xiao-pan, YUAN Xiang-long, WEI Guan-yi, FANG Sui, DENG Hu, LI Zhi-ming. Determination of Minor Isotope Ratios in Uranium by Multiple Collector Inductively Coupled Plasma Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2019, 40(6): 518-524. DOI: 10.7538/zpxb.2018.0167
    [2]WU Jing, CHAO Jing-bo, PENG Yang, WANG Jun, YANG Xiao-jin. Determination of Chromium and Mercury in Sediments by Isotope Dilution-Inductively Coupled Plasma Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2017, 38(6): 647-654. DOI: 10.7538/zpxb.youxian.2016.0066
    [3]HU Xiao-nan, LI Li-li, CHEN Yan, ZHU Liu-chao, CHANG Li. Determination of Pu Isotope Ratio in Environmental Water Samples by IC-MC-ICP-MS[J]. Journal of Chinese Mass Spectrometry Society, 2016, 37(2): 173-179. DOI: 10.7538/zpxb.youxian.2016.0001
    [4]CHEN Xue, WANG Jun, FENG Liu-xing, YANG Xiao-jin. Accurate Determination of K, Ca and Mg in Human Serum by Isotope Dilution Inductively Coupled “Cold” Plasma Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2016, 37(1): 31-36. DOI: 10.7538/zpxb.youxian.2015.0049
    [5]REN Tong-xiang, ZHOU Tao, LU Hai, ZHOU Yuan-jing, WANG Jun. Determination of Silicon Isotope Ratios by Collision-Cell Multi-Collectors Inductively Coupled Plasma Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2015, 36(2): 104-110. DOI: 10.7538/zpxb.youxian.2014.0066
    [6]High-precision Measurement of Hf Isotopic Reference Values for the U-Pb Geochronology Standard Zircons by Multi-collector Inductively Coupled Plasma-Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2005, 26(3): 149-149.
    [7]Determination of Boron in High Purity Quartz by Isotope Dilution-Inductively Coupled Plasma Mass Spectrometry[J]. Journal of Chinese Mass Spectrometry Society, 2004, 25(2): 77-77.
    [8]Study on Accuracy and Precision of Isotope Ratios Measured by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)[J]. Journal of Chinese Mass Spectrometry Society, 2003, 24(3): 441-441.
    [9]Determination of Trace Lead in Seawater Using Isotope Dilution by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)[J]. Journal of Chinese Mass Spectrometry Society, 2003, 24(2): 337-337.
    [10]Determination of Boron in Uranium with ID-ICP-MS after Cation Ion-exchange Separation[J]. Journal of Chinese Mass Spectrometry Society, 2002, 23(2): 65-65.

Catalog

    Article views (2454) PDF downloads (1864) Cited by()

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return