DOU Hong-liang, ZHANG Xiao-hua, ZHENG Qing-lin, WANG Chuan-bo. Determination of Melamine in Milk Powder by GC/MS[J]. Journal of Chinese Mass Spectrometry Society, 2009, 30(增刊): 167-169.
Citation: DOU Hong-liang, ZHANG Xiao-hua, ZHENG Qing-lin, WANG Chuan-bo. Determination of Melamine in Milk Powder by GC/MS[J]. Journal of Chinese Mass Spectrometry Society, 2009, 30(增刊): 167-169.

Determination of Melamine in Milk Powder by GC/MS

  • Melamine in milk power was determined by gas chromatography-mass spectrometer(GC/MS) made in china. The results show that the method has a good liner range between 0.005 mg•L-1 and 2 mg•L-1, R2 is 0.999 7. The limit of detection is 0.001 mg•L-1, the quantitative detection limit is 0.02 mg•kg-1. The recovery for melamine is in the range of 94%-102%. The method has good sensitivity, selectivity and repeatability, which has been applied to the determination of melamine in milk powder.