痕量铅的高精度同位素测量中分馏效应校正技术研究

Calibration Technique of the Fractionation Effect in High Precision Isotopic Ratio Analysis for Trace Lead

  • 摘要: 采用经验校正(K值校正)和双稀释剂校正方法对1~50 ng铅同位素标准样品进行分馏效应的校正效果比较。当样品量大于10 ng时,采用传统的经验校正法,同位素比值与标称值的偏差由0.1%~0.6%减小到0.02%~0.38%;采用双稀释剂校准法,偏差由0.08%~0.45%减小到0.001%~0.14%。当样品量小于10 ng时,两种校准方法未见明显效果,说明影响同位素测量准确度的因素除分馏效应外还存在其它因素。

     

    Abstract: Fractionation effects were corrected by empirical (K correction) and double-dilution methods in isotopic measurements of 1—50 ng lead standard sample. Results showed that the relative deviations of 206Pb/204Pb, 207Pb/204Pb, 208Pb/204Pb could be reduced from 0.1%—0.6% to 0.02%—0.38% for Pb sample more than 10 ng using empirical methods, and using double-dilution methods it could be reduced from 0.08%—0.45% to 0.001%—0.14%. But no obvious effect were detected for Pb sample less than 10 ng of two methods, which implied other facetors that affceted the precision besides fractionation effects.

     

/

返回文章
返回