质谱成像空间稳定同位素示踪分析的基质效应考察

Investigation of Matrix Effects in Mass Spectrometry Imaging-Based Spatial Stable Isotope Tracing Analysis

  • 摘要: 基于质谱成像(MSI)技术的空间稳定同位素示踪分析为揭示生物组织代谢异质性和动态变化规律提供了强有力的工具。然而,MSI分析中的基质效应可能影响定量准确性,并降低不同组织之间代谢物水平的可比性,空间稳定同位素示踪分析是否受组织基质效应的影响尚不明确。本研究采用空气动力辅助解吸电喷雾电离质谱成像(AFADESI-MSI)技术,通过组织匀浆条加标、单器官及整体动物切片喷涂加标3种样本制备方式,将13C标记的葡萄糖作为同位素内标添加至组织中或表面,用于评估不同组织基质效应对目标分子离子强度的影响,计算标记率探究空间稳定同位素示踪中的基质效应。结果表明,直接比较MSI分析中代谢物离子信号强度时,不同组织的基质效应导致目标离子强度存在显著差异,且线性相关性较差。然而,经过稳定同位素内标校正后,各组织的线性相关系数和标准曲线的重合度显著提高,有效降低了组织基质效应的干扰。进一步分析不同组织间标记代谢物的标记率差异时,计算得到的标记率不受组织基质效应的影响,表明空间稳定同位素示踪分析中的标记率计算能够确保跨组织的可比性。通过液相色谱-质谱(LC-MS)联用技术进行验证,证实了标记率在不同组织间具有一致性,未受组织基质差异的干扰。本研究为MSI分析、空间分辨代谢组学和体内稳定同位素示踪研究提供了技术参考和方法依据。

     

    Abstract: Mass spectrometry imaging (MSI) combined with stable isotope tracing has emerged as a powerful analytical approach for investigating the spatial heterogeneity and dynamic reprogramming of metabolic networks in biological tissues. This approach enables in situ visualization of metabolite distributions together with isotopic labeling patterns, providing invaluable insights into tissue-specific metabolic activities and inter-organ metabolic crosstalk. However, MSI analysis is inherently susceptible to matrix effects, which originate from variations in tissue composition, ionization efficiency, and microenvironmental factors. For instance, differences in the chemical environment of various tissues can influence the ionization process, leading to biased or inconsistent ion intensities. Therefore, it is critical to evaluate how these matrix effects might affect the accuracy and reliability of stable isotope tracing results, especially within complex biological systems involving multiple tissues and organs. In this study, the influence of tissue matrix effects on MSI-based spatial stable isotope tracing analysis was systematically examined by airflow-assisted desorption electrospray ionization mass spectrometry imaging (AFADESI-MSI). Three distinct sample preparation methods were employed, including spiked tissue homogenate strips, stable isotope spraying on single-organ tissue sections, and isotope spraying on whole-body animal sections. Uniformly 13C-labeled glucose isotopologues (13C2- and 13C6-glucose) were used to evaluate the impact of tissue matrix effects on the ion intensities of target molecules. The isotopic labeling ratios were calculated to assess the matrix effects on the quantification accuracy of stable isotope tracing. The results showed that when the signal intensities of target ions in MSI analysis were directly compared, differences in tissue matrix effects led to significant variations in ion intensities and a poor linear correlation. These discrepancies highlight the challenges associated with comparing metabolite levels across tissues. However, after the application of stable isotope internal standard correction, the linear correlation coefficients and consistency of calibration curves were significantly improved, effectively reducing interference from tissue matrix effects. This enhancement in correlation accuracy allows for more reliable comparisons between tissue types. Further analyses revealed that the calculated labeling fractions across distinct tissues were unaffected by matrix effects, indicating that the determination of labeling fractions in spatial stable isotope tracing ensured cross-tissue comparability. Verification through liquid chromatography-mass spectrometry (LC-MS) confirmed that the calculated labeling fractions were consistent across tissues and not influenced by matrix differences, further supporting the robustness of the developed method. This study provides important technical references and methodological foundations for MSI analysis, spatially resolved metabolomics, and in vivo stable isotope tracing studies, offering a more reliable analytical strategy for investigating complex biological systems and metabolic processes.

     

/

返回文章
返回