国产高分辨飞行时间质谱仪在药物分子结构鉴定中的应用

Analysis of Pharmaceutical Molecules for Structure Identification by Domestic High Resolution Time-of-Flight Mass Spectrometer

  • 摘要: 为了考察自主研发的API-TOF MS 10000电喷雾高分辨飞行时间质谱仪(ESI-TOF MS)对离子准确质量数测定的准确性及其在新药研发中的应用前景,采用该仪器对新药研发中涉及的291个天然产物及合成物进行质谱分析。以三氟乙酸钠为内标物进行仪器质量轴校准,将测试得到的目标离子准确质量数与理论准确质量数对比,计算两者的相对误差。结果表明:248个样品的分子质量测试结果与预期一致;对其中202个样品进行高分辨质谱测试,所得实验值与理论准确质量数之间的相对误差均小于5×10-6;与预期不一致的实验结果提示样品的结构解析有误,并得到了其他波谱数据的验证。采用API-TOF MS 10000和美国AB Sciex公司的QSIAR Elite型Q/TOF MS对抗癌药物马来酸阿法替尼和降糖药利格列汀杂质进行质谱分析。结果表明,自主研发的API-TOF MS 10000与国际同类仪器对药物分子准确质量数的测试结果无明显差异,可以提供化合物准确的分子质量和分子式信息,是新药研发中药物分子结构解析的有力工具。

     

    Abstract: Electrospray ionization time-of-flight mass spectrometry (ESI-TOF MS) is one of the important tools for structure identification of molecules in drug research and development, in particular, the accuracy mass measured by high-resolution mass spectrometer can provide the information regard to elemental composition for the test compound. In recent years, with the rapid development of pharmaceutical industry, the demand for high-resolution mass spectrometer is increasing in the field of pharmaceutical research and development. The self-developed high-resolution time-of-flight mass spectrometer API-TOF MS 10000 has been designed in Guangzhou on 2011. The mass resolution of 10 000 can be achieved and the instrument detection limit is less than 2.5 μg/L (reserpine), mass ranges cover 30-5 000 Da. In order to verify the performance of this equipment for structure identification and the determining accurate mass of drug molecules, 291 natural products and synthesis products were analyzed on API-TOF MS 10000. Sodium trifluoroacetate was used as internal standard substance to calibrate the instrument, and the relative mass error was obtained by comparing the measured and theoretical mass. The molecular weights of 248 compounds determined by the machine are accord with their identified results, which have a large range of molecular weight from 145 to 5 826 Da. And the 248 compounds cover most of the chemical structure of natural products, such as flavonoids, alkaloids, lignins, phenylpropanoids, coumarins and so on. Accurate mass measurements were performed to confirm 202 compounds’ formula. The results reveal that all the relative mass error is less than 5×10-6, 76.7% of the relative mass error is less than 2.5×10-6. Other results (43 compounds) were inconsistent with the expected structure indicating the original proposed structure was wrong and the other spectra were also provided information to demonstrate the point. In conclusion, the domestic time-of-flight mass spectrometer can accurately determine the mass of drug molecules and isotope abundance ratio, and it has been demonstrated as a powerful tool to provide the molecule and formula information for structure elucidation of pharmaceutical molecules.

     

/

返回文章
返回