Abstract:
A method for the determination of trace arsenic in high purity cobalt with hydride generation-ICP-MS was reported. Due to the interference of the oxide complex ions
59Co
16O
+, the background concentration of
75As is about 100 µg•L
-1 in 1 g•L
-1 cobalt solution, which lead to the impossibility of the direct analysis for the trace As in high purity cobalt. However, the As background concentration can be decreased to several µg•L
-1 if the simple HG system was applied and the 0.05 µg•g
-1 of determination limits can obtained with this method.