Abstract: Double focusing inductively coupled plasma spectrometry (ICP-MS) combing with microconcentricnebulizer have been first applied to the simultaneous determination of Li, Cd, Cs, Pb, Sb, Ba, Tl, Bi ,U, Al, Fe, Mn, Co, Cu and Zn in snow and ice at the 1 pg/ mL and 100 ng/mL level in Tibetan plateau. The volume percentage of concentric HNO\-3(60% HNO\-3) in samples was investigated and 1% was the appropriate percentage in samples considering the samples characteristics in Tibetan plateau. Measurement repeatability based on a snow sample in terms of RSD varies
from 2.5% to 17.5% depending on the elements. The accuracy of the analytical method was verified by analysis of a certified reference material (SLRS-4, riverine water). -
-
[1] Barbante C, Cozzi G, Capodaglio G, et al. Trace Element Determination in Alpine Snow and Ice by Double Focusing Inductively Coupled Plasma Mass Spectrometry with Microconcentric Nebulization[J]. Journal of Analytical Atomic Spectrometry, 1999,14:1 413-1 438. [2] 李月芳, 姚檀栋, 王宁练,等. 电感藕合等离子体质谱仪直接测定冰芯样品中的痕量铅[J]. 岩矿测试, 2001, 20(1): 11-14. [3] 李月芳, 唐富荣. ICP-MS法直接测定冰芯样品中超痕量镉[J]. 分析实验室,2001, 20(3): 25-27. [4] Li Y,Yao T,Li Zh,et al. Recent changes of Heavy metals in a high elevation ice core from Muztag Ata in east Pamirs[J]. [5] Annals of Glaciology, 2006,in press.