电感耦合等离子体质谱法测定乳制品中碘含量的方法研究

Determination of Iodine in Dairy Products by ICP-MS

  • 摘要: 采用四甲基氢氧化铵和过氧化氢提取样品,碲(Te128)作内标,电感耦合等离子体质谱法测定乳制品中碘含量。方法的检出限为0.09 μg•L-1,线性范围为0~400 μg•L-1,相关系数r优于0.999 5。利用不同浓度的相对标准偏差考察方法的精密度,RSD%均在5.0%以下;采用不同浓度的加标回收率考察方法的准确性,加标回收率均在89.3%~116.5%之间;美国NIST(1549)标准物质的测定结果与标实值一致。实验结果表明所建立的方法灵敏度高、简便、快速、准确,适用于乳制品中碘含量的测定。

     

    Abstract: The samples were extracted by tetramethylammonium hydroxide and hydrogen peroxide. Using Te128 as the internal standard, the iodine contents in dairy products were determined by inductively coupled plasma mass spectrometry(ICP-MS). Detection limit of the method is 0.09 μg•L-1, while the linear range is 0—400 μg•L-1 and the linear coefficient(r) is better than 0.999 5. The precision of the method (RSD%) are all less than 5%. Precision and accuracy are evaluated by analysis of certified reference materials NIST 1549 non-fat milk powder. The determined iodine contents were in good agreement with the certified values. The recoveries of real samples are from 89.3%to 116.5%. The experiments show that this method is sensitive, simple and rapid, and accurate for the determination of iodine contents in dairy products.

     

/