Abstract:
In order to analyze uranium isotopic ratios accurately by mass spectrometry, reference materials are used to correct the mass bias and detector bias. Isotopic ratios of UTB U Series working reference material which have been developed in our country last century were measured by multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS). In this report, the isotopic ratios were corrected by measuring CRM U series reference material. Hydride rate UH
+/U
+ of different sample introduction devices were studied. The linearity of the Daly detector was also investigated. Preliminary results have indicated that there is no significant difference between the measurement value and reference value.